-
Different brands of blood glucose meters may use different technologies, so the data results are not comparable. According to the international regulations, the difference between the results measured by the glucose meter and the laboratory at the same time in the same sample shall not be more than 20%. If the difference exceeds the above range, it indicates that the glucose meter itself may have quality problems. The problem is not the instrument in most cases, but other reasons (specimen problem, improper operation, expired test strip, etc).
-
From the strips vial to find a separate password card in blue packaging, in the condition of instrument shutdown, insert into the instrument and the screen will display three numbers. Confirm with the correction code on the test vial, it should be consistent and display “OK”. Pull out the password card. In the condition of shutdown, take a test strip insert instrument. Waiting for the screen to display a strip pattern and a little dot flashing. Then you can test and the result will display automatically in 5 seconds.
-
On the one hand, if the test strip is exposed to the air for a long time, it can absorb the moisture in the air and result in the value is inaccurate. On the other hand, form a habit of covering immediately after using. It can prevent other things interference to forget to cover that lead to the test strips of whole vial waste finally.
-
Fasting blood glucose measured with a glucose meter tends to be lower than that measured by a clinical laboratory, with a difference of 12~20%. It should be noted that in the postprandial state(such as 2 hours after a meal), the results of the glucose meter were similar to those of the clinical laboratory. Suggestion: If you doubt the accuracy of the glucose meter, when you take venous blood to test blood glucose(plasma blood glucose), at the same time, you can test the finger blood glucose(whole blood glucose). Then compare the two values, if the difference is less than 20%, indicating that the glucose meter is relatively accurate.
-
We know that the body's blood glucose is affected by many factors (such as diet, exercise, sleep, mood, etc.) and always in constant changes, sometimes even fluctuating greatly. In the same fasting state, blood glucose values at 6 a.m. and 9 a.m. are different. Suggestion: When comparing blood glucose values, take the blood sample at the same time as possible. Fasting blood glucose: After fasting and rest for 8~10 hours, blood samples were collected before 8 a.m. the next day. Postprandial 2 hours blood glucose: refers to the 2 hours calculated from the first bite of the meal.
-
Sleep disorders and mood changes such as nervousness, anxiety, anger, panic and painful stimuli can cause blood glucose to fluctuate. Suggestion: When analyzing and evaluating test results, we must take all the factors that affect blood glucose into consideration. Maintain emotional stability before testing your blood glucose.
-
①Whether the battery is installed (confirm “yes” to enter the next step)
②Whether the battery is installed correctly (confirm “yes” to enter the next step)
③Open the “on-off button”, whether the screen is displayed (confirm “no” to enter the next step)
④Whether the battery is out of power (confirm “no” to enter the next step)
⑤Contact customer service to replace an instrument.
-
Reason: Blood sampling before the blood drop symbol to display
Solution: Wait for the blood drop symbol to display before blood sampling
-
Reason: Use the used blood glucose test strip or the test strip is exposed to the air for a long time
Solution: Re-test with a new test strip
-
Reason: Operating temperature out of range
Solution: Please measure within the operating range of 10~40℃
-
Reason: Low battery
Solution: Please replace the battery in time
-
Reason: Siphon blood is lacking
Solution: Insert a new test strip and siphon the blood sample to fill the entire window